Abstract : Measurement uncertainty is very important in the inspection work. ISO and other international organizations have already requested it. This paper analyzes the error sources of ultrasonic wave detection longitudinal wave detection and evaluates the uncertainty during the flaw detection process.

Key words : ultrasonic flaw detection longitudinal wave method measurement uncertainty analysis

CLC number : TB553

Foreword

Measurement uncertainty (referred to as uncertainty) is used to characterize the dispersion of reasonable measurement values. It is an important parameter to evaluate measurement results [1 , 2] , which is the correct evaluation and use of standardized measurement results, international standardization, etc. In 1993, the organization jointly developed the "Guidelines for the Expression of Measurement Uncertainty" and applied it widely. In 1996, the domestic uncertainty in the measurement and verification work began to be promoted, but it is not widely used in the quantitative analysis of non-destructive testing. In order to facilitate international integration, this paper applies uncertainty to non-destructive testing and discusses it with you.

The error of longitudinal wave flaw detection comes from the flaw detector, probe, test block, workpiece, coupling agent, etc. There are measurement uncertainties in the process of defect location measurement and quantitative analysis. The evaluation uncertainty can indicate the results of flaw detection analysis. Reliability, has a good promotion value.

1 Location of defects

The longitudinal wave scanning speed is adjusted by 1:n, and the horizontal scale value of the front edge of the defect wave is Ï„ f , and the distance from the defect to the probe is [3] :

X f =nτ f

(1)

1.1 Instruments, workpieces

(1) EPOCH-II ultrasonic flaw detector; (2) 2.25P0.75' probe; (3) CSK-I test block; (4) workpiece with a thickness of 100 mm.

1.2 Determination of position

The instrument adjusts the scanning speed by 1:1, the horizontal scale value is 69.64, and the defect position:

X f =nτ f =69.64 mm

1.3 Uncertainty evaluation

(1) Adjusting the uncertainty caused by the scanning speed When the scanning speed is adjusted, a reading error due to the minimum scale limit of the display screen is generated. The minimum level of the instrument used is 0.1, and the reading can reach the minimum scale of 1/10, that is, the reading error distribution range is (-1/10, +1/10) scale, which is uniform distribution, standard uncertainty (with relative uncertainty) Representation) [4] :
31-1.gif (646 bytes) ; its unreliability is 50%, degree of freedom v 1 = 0-1.gif (119 bytes) (50%) -2 = 2
(2) The level affects the horizontal linearity of the instrument affecting the positioning accuracy of the defect. The horizontal linearity of the instrument is 5.7×10 -3 , subject to the normal distribution (k=3), and its standard uncertainty:
31-2.gif (551 bytes) (Class B): its value is reliable, the degree of freedom v 2 =∞
The horizontal scale error is affected by the resolution of the instrument (resolution 0.01), subject to uniform distribution, the error is 1/2 resolution, the distribution range (-0.005, +0.005) [5] , its standard uncertainty (relatively uncertain) Degree representation):
31-3.gif (703 bytes) ; (Class B); reading reliability is 50%, degree of freedom v 3 = 0-1.gif (119 bytes) (50%) -2 = 2
The horizontal scale values ​​were measured 10 times, with an average of 69.64 and a standard deviation of s=0.21. The standard uncertainty (represented by relative certainty):
31-5.gif (585 bytes) ; (Class A): Degree of freedom v 4 = n-1 = 9
The horizontal scale synthesis standard uncertainty is the synthesis of the three standard uncertainties of the instrument's horizontal linearity, instrument resolution, and horizontal scale value repeated observation, which is the synthesis of the above three Class A and Class B:

31-6.gif (918 bytes)

(3) The results are used to evaluate the transfer function of each component: 32-1.gif (721 bytes) =n=1
The location of the defect, its synthetic uncertainty:

32-2.gif (1532 bytes)

Effective degrees of freedom:

32-3.gif (836 bytes)

Take the confidence level 95%, t P (0.95)=1.96
The defect position is: X f (1±t P u C (X f ))=69.64 (1±8.8×10 -3 )
Its absolute uncertainty (synthetic standard) u C (X f )=X f . u C (X f )=0.3127

2 Quantitative analysis of defects

The defect equivalent size is calculated by the following formula:

32-4.gif (497 bytes)

(2)

It can be seen from the formula (2) that the defect equivalent size is affected by the defect position, the thickness of the workpiece, and the amount of attenuation.

2.1 Change in attenuation

The variation in attenuation during detection is caused by vertical linearity, probe frequency, attenuator error, couplant, medium attenuation, and the like.

2.1.1 The effect of vertical linearity is obtained by the following equation: vertical linearity and attenuation:

32-5.gif (858 bytes)

(3)

Where: H 0 —— ideal wave height H 1 —— measured wave height (H 1 1 is replaced by H 1 1 for one measured wave height, H′′ 1 is substituted for H 1 for the second measured wave height)
1 wave height effect transfer function 32-6.gif (1054 bytes) ×0.4342944=10.9
The linearity error is 4.7×10 -2 , subject to a normal distribution (k=3), and the standard uncertainty (represented by relative uncertainty):
32-7.gif (575 bytes) (Class B); its value is stable, and the degree of freedom v 5 =∞.
The reading error is 1/2 of the resolution, subject to a uniform distribution range (-0.5, +0.5), and standard uncertainty (expressed in relative uncertainty):
32-8.gif (606 bytes) ; (Class B) its estimated reliability is 50%, degree of freedom v 6 = 0-1.gif (119 bytes) (50%) -2 = 2
The uncertainty of a wave height is synthesized by the above two types of Class A and Class B, namely:

32-9.gif (811 bytes) 10 -2

2 secondary wave height effect transfer function 32-10.gif (981 bytes)
The second wave height has a linearity error of 4.7×10 -2 , and its standard uncertainty is:
u 7 =u 5 =1.57×10 -2 (Class B); degrees of freedom v 7 =v 5 =∞
The reading error is the same wave, u 8 =u 6 , v 8 =v 6

32-11.gif (673 bytes)

The uncertainty caused by the linear error is the synthesis of the primary wave height and the second wave height standard uncertainty, namely:

32-12.gif (1808 bytes)

2.1.2 Effect of frequency-induced attenuation frequency on attenuation: 32-13.gif (633 bytes)
Transfer Function: 32-14.gif (681 bytes)32-15.gif (1134 bytes)
Frequency error 10%, subject to normal distribution (k=3), standard uncertainty:
32-16.gif (453 bytes) ; its value is stable, the degree of freedom v 9 =∞
Standard uncertainty that causes attenuation:

33-1.gif (1056 bytes)

2.1.2 Influence of coupling agent

Change the thickness of the coupling agent, measure its attenuation, and analyze its uncertainty by statistical method (the number of times of this test is 10). The average value of the test is 5.5 dB, the standard deviation is s=0.566, and the standard uncertainty is:

33-2.gif (633 bytes)

Expressed in relative uncertainty: 33-3.gif (515 bytes) 10 -2 ; degrees of freedom v 10 = 10-1=9
The error caused by the reading is 1/2 of the resolution, subject to uniform distribution, and its distribution range is (-0.05, +0.05), standard uncertainty:

33-4.gif (732 bytes)

Expressed in relative uncertainty: 33-5.gif (581 bytes) ; reading reliability 50%, degree of freedom v 11 = 0-1.gif (119 bytes) (50%) -2 = 2
Effect of attenuator accuracy (measured at 12dB):
Attenuator error 0.5dB, subject to normal distribution (k=3), standard uncertainty:

33-6.gif (656 bytes)

Expressed in relative uncertainty 33-7.gif (496 bytes) 10 -2 ; its value is stable, and the degree of freedom v 12 =∞.
The uncertainty of the amount of attenuation caused by the couplant is the synthesis of the above three Class A and Class B:

33-8.gif (1067 bytes)

2.1.4 Media attenuation Repeat inspection of the workpiece (10 times), the medium attenuation, the mean value is 5.3, the standard deviation s=0.2 dB, the standard uncertainty is determined:

33-9.gif (645 bytes)

Expressed in relative uncertainty: 33-10.gif (598 bytes) ;degree of freedom v 13 =10-1=9
Standard uncertainty due to reading: u' 14 = u 11 = 0.029 dB (Class B)
Expressed in relative uncertainty: 33-11.gif (534 bytes) 10 -3 ; degrees of freedom v 14 =v 11 =2
Effect of the attenuator: u 15 =u 12 =0.167 dB (Class B)
Expressed in relative uncertainty: 33-12.gif (499 bytes) ×10 -2 ; degree of freedom v 15 =v 12 =∞.
The uncertainty of medium attenuation is the synthesis of the above three Class A and Class B: 33-13.gif (549 bytes)
33-14.gif (591 bytes)
2.1.5 Attenuation Amount Changes of repeated observations The workpiece is measured 10 times, with an average of 6.2 and a standard deviation of s=0.58. The standard uncertainty is:

33-15.gif (643 bytes)

Expressed in relative uncertainty: 33-16.gif (617 bytes) ;degree of freedom v 16 =10-1=9
Standard uncertainty due to reading error: u 17 =u 11 =0.029 dB (Class B)
Expressed in relative uncertainty: 33-17.gif (542 bytes) 10 -3 ; degrees of freedom v 17 =v 11 =2
The effect of the attenuator, its standard uncertainty: u 18 =u 12 =0.167 dB (Class B)
Expressed in relative uncertainty: 33-18.gif (510 bytes) 10 -2 ; its value is stable, and the degree of freedom v 18 =∞.
Synthetic standard uncertainty:

33-19.gif (1118 bytes)

At this point, the uncertainty of the various factors affecting the attenuation is:

33-20.gif (1890 bytes)

2.2 Thickness effect

The average value of 100 measurements was 100, the standard deviation s=0.15 mm, standard uncertainty:

33-21.gif (828 bytes)

Expressed in relative uncertainty: u' 19 = 4.74 × -4 ; degree of freedom v 19 = 10-1 = 9
The caliper minimum reading is 0.02, the error is subject to uniform distribution, the distribution range (-0.01, +0.01), its standard uncertainty:

33-22.gif (775 bytes)

Expressed in relative uncertainty: u' 20 = 5.8 × 10 -5 ; degree of freedom v 20 = 0-1.gif (119 bytes) (50%) -2 = 2
The caliper uncertainty is 0.05 (k=3), subject to normal distribution, standard uncertainty:

34-1.gif (772 bytes)

Expressed in relative uncertainty: 34-2.gif (578 bytes) 10 -4 ; its value is stable, and the degree of freedom v 21 =∞.
The standard uncertainty of workpiece thickness synthesis is synthesized by the above three items, namely:

34-3.gif (1122 bytes)

2.3 defect equivalent size

34-4.gif (1119 bytes)

Transfer Function:

34-5.gif (2389 bytes)

Synthetic standard uncertainty:

34-6.gif (2157 bytes)

Expressed in relative uncertainty: u' c (Φ) = 3.08 × 10 -2
Effective degrees of freedom:

34-7.gif (926 bytes)

The confidence level is 95%, then the coverage factor: t P (4) = 2.78

U=t P . u c (Φ f )=2.78×0.0613=0.17

The final result of the equivalent size is Φ f =1.99±0.17 mm (0.17 mm is the measurement uncertainty of the test results)

3 Conclusion

Through the calculation and evaluation of the measurement uncertainty of the quantitative analysis of longitudinal wave flaw detection, the reliability degree of the quantitative analysis results of the flaw detection can be known, and the error source of the test process and the result analysis and the influence degree of the joint test result can be determined. The evaluation of measurement uncertainty is an important part of the test results and can be applied to various fields of quantitative analysis of non-destructive testing.

About the author: Wu Jianping, male, born in 1962, engineer
Author:
Wu Jianping (Jiangling Motor Co., Ltd., Nanchang 330001, China)
Wei Zhihua (Jiangling Motor Co., Ltd., Nanchang 330001, China)

references

[1] BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML, Guide to the Expression of Uncertainty in Measurement, ISO, 1995;
[2] Xia Jizhen, Acoustic Nondestructive Testing Technology, Nanchang: Nanchang Institute of Aeronautical Industry, 1991;
[3] Liu Zhimin, Evaluation of Measurement Uncertainty, Beijing: China Institute of Technical Supervision and Information, 1995;
[4] Editor-in-Chief Chen Yuqin, Application Example of Measurement Uncertainty "93 International Guide", Beijing: China Metrology Press, 1998.

Google_ad_client = "pub-9310853799989062";google_ad_width = 728;google_ad_height = 90;google_ad_format = "728x90_as";google_ad_type = "text";google_ad_channel = "9247596413";google_color_border = "FFFFFF";google_color_bg = "FFFFFF";google_color_link = "000000 ";google_color_text = "000000";google_color_url = "000000";

Crimped Wire Mesh  
Precrimped Wire Mesh 1.0-13.5mm 
square openg,strong struction. 
manufacuter ISO 9001

Product:Crimped Wire Mesh 
Material: 302, 304, 304L,309,309S,310,310S 316, 316L.321.329,430,nickel,monel ,carbon steel, Aluminum ect.
Characteristics: strong structure, durability and mesh well-distributed..
Usage:Mining, coal industry,construction, chemical and fertilize.
 

Crimped Wire Mesh

Crimped Wire Mesh,Crimped Mesh,Stainless Steel Crimped Wire Mesh,Double Crimped Wire Mesh

CSTAR WIRE MESH ENTERPRISES CO., LTD. ,